Heidenhain IK 5000 Addendum Manual de usuario Pagina 49

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 100
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 48
IK 5000 QUADRA-CHEK 49
1.1 Operation
Perform Profile3D fit analysis
When all the required points have been probed, the points can be fit
to the nominal part model.
When the fit operation is complete, the green probed points will be
shifted to match the part model as closely as possible and form error
whiskers will indicate differences between the probed locations and
the nominal part surfaces.
To perform a fit analysis:
Click Fit
Points shown before fit calculation Form error whiskers shown after fit operation
To decrease the fit analysis time, well distributed points all
around the part under measure are recommended. In
addition, part alignment and datum can be performed
before commencing the Profile3D measurement or using
the Six Point alignment function in the Profile3D
measurement.